Rezeroing a transducer – ZOLL Propaq M Rev B User Manual

Page 122

Advertising
background image

C

HAPTER

11 M

ONITORING

I

NVASIVE

P

RESSURES

(IBP)

11-4

www.zoll.com

9650-0803-01 Rev. B

Figure 11-2 IBP Channel Parameter Control Pane

6.

Select Zero Probe. The unit displays the message

ZEROING

in the IBP Channel’s numeric

display.

7.

The message ZEROED appears in the IBP channel’s numeric display.

8.

Close the transducer’s stopcock.

9.

If the unit was unable to zero the transducer, the message ZERO REJECTED appears in the
IBP channel’s numeric display. The unit will not display pressure values for the IBP channel
until it has successfully zeroed the transducer and established an acceptable zero reference.

Check that the unit is open to atmospheric air and that it is properly connected to the unit, then
try zeroing the transducer again. The Propaq M unit will not zero the transducer if it detects
pulsation in the pressure channel, if there is too much noise in the signal, or if transducer’s
offset is too great.

If you cannot zero the transducer after several attempts, replace the transducer or the transducer
cable.

Rezeroing a Transducer

You can re-zero a transducer at any time by opening the transducer stopcock to atmospheric air.

If the unit accepts the new zero reference value, the unit displays pressure values that are based
on that value and adjusts the waveform according to the new scale.

Warning!

If you select an IBP channel after the channel has been successfully zeroed and is
currently monitoring a pressure waveform, the unit will display the message, ZERO
REJECTED

, in the IBP channel’s numeric window. This message will preempt the

display of the valid invasive pressure numerics.

P1

Lower

Upper

P1 Systolic Alarm

90

150

Source Label

P1

Display Format

S/D (M)

P1 Diastolic Alarm

60

90

P1 Mean Alarm

70

120

Zero Probe

Advertising