Cleaning and Maintenance

Monitoring equipment must be maintained to be ready for immediate use. To ensure the
readiness and optimum working condition of the Propaq M unit, you should perform the
cleaning and maintenance instructions recommended in this chapter.

Inspection and Cleaning instructions

Before cleaning the monitor, battery, AC power adapter, or any accessories, thoroughly inspect
them:

Look for any signs of damage and any improper mechanical function of buttons or
connectors.

Gently bend and flex cables, inspecting them for damage or extreme wear, exposed wires, or
bent connectors.

Confirm that all connectors engage securely.

Immediately report any sign of damage or malfunction to your service department.

At least once per year thoroughly inspect the battery and the AC adapter power cord for
damage or extreme wear. In addition to this inspection, authorized personnel should complete
performance and calibration testing at regularly scheduled intervals, which should not exceed
one year.