Using the resp parameter control panel – ZOLL Propaq M Rev B User Manual

Page 78

Advertising
background image

Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)

7-4

www.zoll.com

9650-0803-01 Rev. B

Respiration Rate Alarm Limits

Initially, the Resp Alarm Settings menu specifies that Resp alarms are enabled (

ON

)

or

disabled (

OFF

),

and displays the default Upper and Lower respiration rate alarm limits. The

upper and lower limits can be

ON

or

OFF

(default is

OFF

).The following table lists the default

respiration rate alarm limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:

Using the Resp Parameter Control Panel

To configure alarms through the

Resp

Parameter Control Panel, use the navigation keys to

highlight and select the Respiration Rate meter and display the

Resp

Parameter Control Panel:

The

Resp

Parameter Control Panel allows you to set the following parameters:

RR/BR Alarm

-- enable/disable the Resp alarm and set high/low alarm limits.

No Breath Alarm

-- sets the duration of the No Breath alarm.

Resp Monitoring--

enable/disable respiration monitoring.

Resp Lead

-- selects the Resp lead, Ld I (RA-LA) or Ld II (RA-LL), from which the Propaq

M unit calculates the respiration rate. Resp Lead selection is independent of ECG Lead
selection.

CO2/Resp Sweep Speed --

sets the respiratory sweep speed on the display.

Patient Type

Respiration Rate Default

Respiration Rate Range

Adult

Lower: 3 BPM

Upper: 50 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

Pediatric

Lower: 3 BPM

Upper: 50 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

Neonate

Lower: 12 BPM

Upper: 80 BPM

Lower: 0 to 145 BPM

Upper: 5 to 150 BPM

Resp

Lower

Upper

RR/BR Alarm

10

30

No Breath Alarm

30 sec

Resp Monitoring

On

Resp Lead

Ld II (RA-LL)

CO2/Resp Sweep Speed

12.5 mm/s