ZOLL Propaq M Rev B User Manual
Page 63
Advertising
Alarm Options
9650-0803-01 Rev. B
Propaq M Operator’s Guide
5-9
SpO2
(%)
Entire range
Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)
Limit = Numeric – 5
EtCO2
(mmHg)
Entire range
Limit = Numeric + 10
Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.
FiCO2
(mmHg)
Entire range
Limit = Numeric + 5
N/A
Temp
(°C)
Entire range
Limit = Numeric + 0.5
Limit = Numeric – 0.5
Temp
(°F)
Entire range
Limit = Numeric + 0.9
Limit = Numeric – 0.9
Advertising