Alarm Options

9650-0803-01 Rev. B

Propaq M Operator’s Guide

5-9

SpO2

(%)

Entire range

Limit = 100 (Adult and
Pediatric)
Limit = Numeric + 5 (Neo-
nate)

Limit = Numeric – 5

EtCO2

(mmHg)

Entire range

Limit = Numeric + 10

Limit = Numeric - 5 mmHg
unless the numeric falls
below the lower alarm limit
range, in which case Stat
Set sets the lower limit to
15 mmHg.

FiCO2

(mmHg)

Entire range

Limit = Numeric + 5

N/A

Temp

(°C)

Entire range

Limit = Numeric + 0.5

Limit = Numeric – 0.5

Temp

(°F)

Entire range

Limit = Numeric + 0.9

Limit = Numeric – 0.9