Propaq M Product Functions

9650-0803-01 Rev. B

Propaq M Operator’s Guide

1-9

Ready For Use (RFU) Indicator

The Propaq M has an RFU indicator on the front panel that indicates if the device is ready for
use. The RFU indicator has three states which are described in the following table.

State

Description

Action

Ready for Use

The device is ready for use. Patient
monitoring is functional and the bat-
tery is above the low battery capac-
ity.

None required.

Flashing

One or more of the following has
occurred:

The battery is not properly
installed.

A low battery is installed.

A battery fault has occurred.

There is no battery installed
while connected to auxiliary
power.

One or more patient monitoring
parameters have failed self-test
(NIBP, SpO2, CO2, IBP, or
Temp).

The front panel button self-test
failed.

The speech database self-test
failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator contin-
ues to flash, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.

Do Not Use

One or more of the following has
occurred:

The battery is not properly
installed.

No battery is installed and auxil-
iary power is not present.

A very low battery (below soft-
ware shutdown limit) was
installed.

ECG or other critical self-tests
have failed.

Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator contin-
ues to display the Do Not Use sym-
bol, remove the unit from service
and contact the appropriate techni-
cal personnel or the ZOLL Techni-
cal Service Department.