System messages – ZOLL Propaq M Rev B User Manual

Page 117

Advertising
background image

System Messages

9650-0803-01 Rev. B

Propaq M Operator’s Guide

10-11

Setting the CO

2

Sweep Speed

The EtCO

2

sweep speed determines the X-axis scale of the capnogram. For patients with

slower respiration rates, a slower sweep speed will make the capnogram easier to view. You can
specify sweep speeds of 3.13, 6.25, and 12.5 mm/second. The default sweep speed is
6.25 mm/second.

System Messages

When monitoring CO

2

, the Propaq M unit may display the following messages:

System Message

Cause

INITIALIZING

The unit is initializing the CO

2

monitoring function

measurement, is functioning normally, and will display
CO

2

measurement after a brief delay.

CHECK FILTERLINE

The sampling line is not connected. Check
sampling line connection. Replace airway adaptor
or cannula, if defective.

FILTERLINE OCCLUSION

The sampling or exhaust line is blocked. Check
sampling and exhaust lines. Make sure that the
sampling line and any inputs to the patient
breathing apparatus are not connected to the
exhaust port.

If the current FilterLine is correctly

attached, replace the FilterLine.

AUTO ZERO

The Propaq M unit displays AUTO ZERO when the
CO

2

module performs periodic self-maintenance.

During self-maintenance, the CO

2

module performs

one or more of the following tests: ambient pressure
measurement, auto zero, and a flow test. The CO

2

module completes its self-maintenance tests in
approximately 10 seconds.

PURGING

The CO

2

module performs a purging whenever it

detects an occlusion in the line or airway adaptor,
and displays PURGING while it tries to clear the
occlusion. If the CO

2

module cannot clear the

occlusion within 30 seconds, the Propaq M unit
issues an equipment alert and displays the
message FILTERLINE OCCLUSION.

At this point, if

the current FilterLine is correctly attached, replace the
FilterLine.