Chapter 11, Monitoring invasive pressures (ibp), Invasive pressure transducers – ZOLL Propaq M Rev B User Manual

Page 119

Advertising
background image

9650-0803-01 Rev. B

11-1

Chapter 11

Monitoring Invasive Pressures (IBP)

The Propaq M IBP inputs are Type CF defibrillator proof.

This chapter describes how to use the Propaq M unit to monitor invasive pressures (IBP).

The Propaq M unit has three invasive pressure channels: P1, P2, and P3. You can use these
channels to measure arterial, venous, or intracranial pressures using invasive transducers with
5uV/V/mmHg sensitivity. Each channel requires its own connector, cabling, and pressure
transducer.

Invasive Pressure Transducers

The Propaq M unit is compatible with many types of invasive pressure transducers, including
non disposable, disposable dome, and fully disposable transducers. See Appendix B,
Accessories, for a list of compatible transducers. Do not use light-sensitive disposable
transducers.

Use the invasive pressure transducers according to your organization’s established clinical
protocol and follow the manufacturer’s recommendations. Always refer the manufacturer’s
Directions for Use before using a transducer.