Battery

Although the device can operate with auxiliary power alone, ZOLL strongly recommends that
you operate the unit with a battery installed at all times. Operating the unit with a battery
provides a backup in case of ac power shortage. The battery can be automatically recharged
while it is installed in the unit. Keep a fully charged spare battery pack with the monitor at all
times.

Test battery packs regularly. A battery that does not pass the ZOLL charger’s capacity test
might cause the Propaq M unit to shut down unexpectedly.

If the Low Battery indication occurs at any time during operation, immediately replace the
battery pack.

If the LOW BATTERY icon appears, plug the Propaq M unit into a power source or install a
fully charged battery pack. When the warning low battery shutdown prompt appears,
immediately replace the battery pack with a fully charged pack or plug the Propaq M unit into a
power source, as unit shut down due to a low battery condition is imminent.

If mistreated, a battery pack might explode. Do not disassemble a battery pack or dispose of it
in fire.

Operator Safety

Do not use the Propaq M in the presence of oxygen-rich atmospheres, flammable anesthetics,
or other flammable agents (such as gasoline). Using the unit in such environments might cause
an explosion.

Do not use the unit in standing water. Electrical safety might be compromised when the
monitor is wet.

The use of accessory equipment that does not comply with the equivalent safety requirements
of the Propaq M could reduce the level of safety of the combined system. When choosing
accessory equipment, consider the following:

Use of the accessory in the patient vicinity.

Evidence that the safety certification of the accessory has been performed in accordance
with the appropriate IEC (EN) 60601-1 and/or IEC (EN) 60601-1-1 harmonized national
standards.

Always check that the equipment functions properly and is in proper condition before use.

Patient Safety

This equipment should be connected to only one patient at a time.

To ensure patient safety, do not place the monitor in any position that might cause it to fall on
the patient.

Use only high-quality ECG electrodes.

Do not use ECG electrodes if the gel is dried, separated, torn or split from the foil; patient burns
may result from using such electrodes.

Check the expiration date on the electrode packaging. Do not use electrodes after their
expiration date.