Monitoring Display Options

9650-0803-01 Rev. B

Propaq M Operator’s Guide

3-3

Capnography (CO

2

)

CO

2

monitoring measures the CO

2

concentration in a patient’s exhaled breath (End Tidal

Carbon Dioxide --EtCO

2

). CO

2

monitoring can also measure a patient’s breath rate and the

CO

2

concentration in the gasses supplied to intubated patients (Fractional Inspired Carbon

Dioxide -- FiCO

2

). Since FiCO

2

represents the amount of CO

2

present during inhalation, it also

serves as an indicator for rebreathing in non-intubated patients. CO

2

monitoring can be used for

both intubated and non-intubated patients.

The EtCO

2

, breath rate, and FiCO

2

measurements appear in a labeled (

EtCO

2

) numeric display.

The EtCO

2

and FiCO

2

measurements can appear as values given in millimeters of mercury

(mmHg). You can also specify that the Propaq M unit display a CO

2

capnogram in the

waveform trace display area.

Pulse Oximetry (SpO

2

)

Pulse Oximetry monitoring measures the oxygen saturation (SpO

2

) of arterial blood at a

peripheral site such as a finger or toe. SpO

2

monitoring determines the ratio of oxygenated

hemoglobin to total hemoglobin in arterial blood and displays this ratio as percent SpO

2

in a

labeled (

SpO

2

) numeric display. You can also specify that the Propaq M unit display an SpO

2

plethysmograph in the waveform trace display area.

Monitoring Display Options

The Propaq M unit gives you great flexibility in how you display a patient’s vital signs

information. By pressing the Display/Home button (

),on the front panel, you can

successively display the patient’s vital signs information in these three windows:

•

Waveform Display window, which initially displays an ECG waveform trace and numeric
displays for each monitoring function.

•

Trends Status window, which displays a report listing vital signs measurements that the
Propaq M unit logs automatically, and the primary ECG waveform trace.

•

Large Numerics Display window, on which large numeric displays of all vital signs
measurements appear.

The Waveform Display window appears when you power on the Propaq M unit. Initially, the
Waveform Display window displays a single ECG waveform trace. All other monitored values
appear in numeric display areas at the bottom of the screen: