Chapter 13 monitoring 12-lead ecg – ZOLL Propaq M Rev B User Manual

Page 135

Advertising
background image

9650-0803-01 Rev. B

Propaq M Operator’s Guide

13-1

Chapter 13

Monitoring 12-Lead ECG

The Propaq M 12-Lead input is Type CF defibrillator proof.

This chapter describes how to use the Propaq M unit to perform 12-Lead ECG Monitoring.

The Propaq M with the 12-lead ECG Monitoring option provides simultaneous acquisition and
storage of 12-lead information.

Warning!

•

12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals
acquired from adult and pediatric patients in the supine, resting position -- always
ensure that the patient is kept motionless during 12-lead ECG signal acquisition

. Use

of the device to acquire ECG signals from moving or shaking patients may produce
noisy signals that are difficult to interpret.

•

Excessive body hair or wet, sweaty skin may interfere with electrode adhesion.
Remove the hair and/or moisture from the area where the electrode is to be attached.

•

Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.

•

Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within a
few seconds.

•

When not in use, cover the patient cable's V-lead connector with the supplied plastic
cap. Failure to do so may result in a shock hazard during defibrillation attempts.