Monitoring 12-Lead ECG

The Propaq M 12-Lead input is Type CF defibrillator proof.

This chapter describes how to use the Propaq M unit to perform 12-Lead ECG Monitoring.

The Propaq M with the 12-lead ECG Monitoring option provides simultaneous acquisition and
storage of 12-lead information.

Warning!

12-Lead ECG monitoring is intended for the recording of 12-lead ECG signals
acquired from adult and pediatric patients in the supine, resting position -- always
ensure that the patient is kept motionless during 12-lead ECG signal acquisition

. Use

of the device to acquire ECG signals from moving or shaking patients may produce
noisy signals that are difficult to interpret.

Excessive body hair or wet, sweaty skin may interfere with electrode adhesion.
Remove the hair and/or moisture from the area where the electrode is to be attached.

Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.

Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within a
few seconds.

When not in use, cover the patient cable's V-lead connector with the supplied plastic
cap. Failure to do so may result in a shock hazard during defibrillation attempts.