Temperature system messages – ZOLL Propaq M Rev B User Manual

Page 133

Advertising
background image

Temperature System Messages

9650-0803-01 Rev. B

Propaq M Operator’s Guide

12-5

You can select one of the following labels for each temperature channel:

If you don’t select a label, the Temperature channels appear with the default labels of T1 and
T2.

Temperature System Messages

The Propaq M unit may display the following messages when monitoring Temperature.

Note:

The temperature function performs a self test when initially powered on and also
performs system tests automatically, every 10 seconds, while this function is active.

Source Label

Description

ART

Arterial temperature probe

CORE

Central body temperature or tympanic membrane temperature
probe

ESOPH

Esophageal temperature probe

RECT

Rectal temperature probe

SKIN

Skin temperature probe (surface application)

VEN

Ventilator airway temperature probe

NASO

Nasopharynx or nasal/oral temperature probe

System Message

Cause

CHECK PROBE

The temperature probe is disconnected. Check probe
and reconnect it.

PROBE FAILURE

The temperature probe is defective. Replace the
temperature probe.

TEMP DISABLED

A system error has occurred. The Propaq M unit
cannot take temperature measurements and should
be serviced.