Electromagnetic Compatibility Guidance and Manufacturer’s Declaration

9650-0803-01 Rev. B

Propaq M Operator’s Guide

A-11

Guidance and Manufacturer’s Declaration – Electromagnetic
(IEC 60601-1-2 Table 202)

The Propaq M unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Propaq M should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test Level Compliance Level

Electromagnetic Environment
– Guidance

Electrostatic
Discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.

Electrical Fast
Transient / Burst
IEC 61000-4-4

± 2 kV for power
supply lines
± 1 kV for input/
output lines

± 2 kV for power
supply lines
± 1 kV for input/
output lines

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge IEC 61000-4-
5

± 1 kV differential
mode
± 2 kV common mode

± 1 kV differential
mode
± 2 kV common mode

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines IEC 61000-4-
11

<5% U

T

(>95% dip in U

T

) for

0.5 cycle
40% U

T

(60% dip in U

T

) for

5 cycles
70% U

T

(30% dip in U

T

) for

25 cycles
<5% U

T

(>95% dip in U

T

) for

5 sec

<5% U

T

(>95% dip in U

T

) for

0.5 cycle
40% U

T

(60% dip in U

T

) for

5 cycles
70% U

T

(30% dip in U

T

) for

25 cycles
<5% U

T

(>95% dip in U

T

) for

5 sec

Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Propaq M unit requires
continued operation during
power mains interruptions, it is
recommended that the Propaq M
unit be powered from an
uninterruptible power supply or a
battery.

Power Frequency
(50/60 Hz) Magnetic
Field IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

Note: U

T

is the AC mains voltage prior to application of the test level.