Entering patient information – ZOLL Propaq M Rev B User Manual

Page 136

Advertising
background image

C

HAPTER

13 M

ONITORING

12-L

EAD

ECG

13-2

www.zoll.com

9650-0803-01 Rev. B

To assure protection against the effects of defibrillator discharge, use only 12-lead
cables supplied by ZOLL Medical Corporation.

Check the operation and integrity of the Propaq M unit and 12-lead cable regularly
by performing the Daily Operational Verification Test.

When attempting to interpret subtle ECG changes (such as ST segments) use only
the Diagnostic frequency response setting. Other frequency response settings may
cause misinterpretation of the patient’s ECG.

Use only ZOLL-approved accessories to ensure the Type CF defibrillator proof
rating of the 12-Lead input.

Implanted pacemakers may cause the heart rate meter to count the pacemaker rate
during incidents of cardiac arrest or other arrhythmias. Carefully observe
pacemaker patients. Check the patient's pulse; do not rely solely on heart rate
meters. Dedicated pacemaker detection circuitry may not detect all implanted
pacemaker spikes. Patient history and physical exam are important in determining
the presence of an implanted pacemaker.

Entering Patient Information

To enter patient information, press the 12-Lead quick access key (

), then the Patient Info

quick access key (

). The screen displays the Patient Info parameter panel, in which you can

enter the patients’s name, age, gender, and identification number:

Figure 13-1 Patient Info Control Panel

The Propaq MD unit uses the name that you enter in the Patient Info panel to label the 12-lead
ECG monitoring snapshots that it saves.

To enter patient information, use the navigation keys to highlight and select a parameter on the
Patient Info panel, then press the Select key.

12

Patient Info

Patient Last Name

Patient Age

45

Patient First Name

Patient Middle Name

Patient Gender

M

Patient ID