This chapter describes how to use the Propaq M unit to perform Non-Invasive Blood Pressure
(NIBP) measurements using an inflatable cuff to measure arterial pressure.

The Propaq M uses the enhanced NIBP features of Welch Allyn’s SureBP and patented
Smartcuf motion-tolerant

technology.

The SureBP monitoring software allows the Propaq M unit to take an NIBP measurement as the
cuff is inflating, which saves time -- the measurement takes about 15 seconds -- and improves
patient comfort.

The Smartcuf monitoring software enables the Propaq MD unit to make accurate NIBP
measurements in the presence of extreme artifact, weak pulses, and some dysrhythmias by
synchronizing the NIBP measurements with the patient’s R-wave.

Warning!

Observe the patient’s limb periodically to ensure that circulation is not impaired for
a prolonged period of time.

Never use the Propaq M to monitor NIBP on one patient while simultaneously
monitoring ECG on another patient.

If a non-invasive blood pressure measurement is suspect, repeat the measurement. If
you are still uncertain about the measurement, use another method.

Do not use NIBP without proper training. Check that the correct patient mode has
been selected to ensure that the initial inflation pressure is set correctly. Patient
movement, very low pulse volume, or vibration from outside sources can influence
the accuracy of blood pressure measurements.