Restoring default values, Procedure – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 15

Advertising
background image

Restoring Default Values

9650-0902-01 Rev. K

9

Restoring Default Values

You can restore all configuration settings to their factory default settings. Only the display
language, if different from the default, does not change.

Procedure

Follow these steps to restore all configuration settings to default values:

If you want to review configuration settings, press the Change Config softkey. See “Accessing
and Changing Configuration Settings” on page 10 for more information. If you want to leave
System Configuration mode, press the Exit Config softkey. See “Leaving Configuration
Mode” on page 11 for more information.

Step

Action

1.

Power on the unit in System Configuration mode as
described in “Entering System Configuration Mode”
on page 2.

2.

Press the Restore Defaults softkey to reconfigure
settings to their default settings.

Once default settings are restored, the unit displays
the DEFAULTS RESTORED and PRESS CHANGE
CONFIG TO VIEW
messages.

System Configuration

Print

Change

Config

Exit

Config

Read
From

Card

Restore

Defaults

System Configuration

Defaults Restored

Press Change Config to View

Print

Change

Config

Exit

Config

Read
From

Card

Restore

Defaults