Chapter 2

Configuration Settings

50

9650-0902-01 Rev. K

Enable Readiness Test Audible Alert (Future Use)

This setting allows you to specify whether the unit emits a beep when the automatic device
check fails.

Values are as follows:

One Step Padz Installed

This setting allows you to specify whether the R Series unit tests the R Series OneStep
electrodes as part of the readiness test check, and which electrodes it tests.

If this setting is set to any of its values (including No or Base), and CPR, MwP or Complete
electrodes are detected, they will be tested and the resulting status (pass/fail) is included in
the readiness result.

If this setting is set to CPR/MwP/Complete, and none of these OneStep electrodes are
connected, or if the cable is not connected to the unit, the readiness result will indicate a
failure (Electrode: Not Connected).

If this setting is set to Base or No, and the cable is correctly connected to a short (Base
electrode package, test port, or paddles in the wells), the unit will perform the readiness test
and not fail based on electrode type.

Values are as follows:

Note:

The OneStep Padz Installed option should be set to Base unless you are using OneStep
electrodes marked Rev. B or above:

8900-0212-XX Rev. B or above for OneStep Pacing
8900-0213-XX Rev. B or above for OneStep CPR
8900-0214-XX Rev. B or above for OneStep Complete

Possible Values:

No

Default:

No

Possible Values:

CPR/MwP/Complete, No, Base

Default:

CPR/MwP/Complete