Entering system configuration mode, Procedure, System configuration – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 8

Advertising
background image

Chapter 1

Introduction

2

9650-0902-01 Rev. K

Entering System Configuration Mode

In order to make any changes to unit configuration, you must first power on the unit in System
Configuration mode.

Procedure

Follow these steps to power on the unit in System Configuration mode:

Step

Action

1.

Turn the selector switch to OFF, and wait 10
seconds.

Simultaneously press and hold the right-most and
left-most softkeys (below the monitor) while turning
the selector switch to MONITOR (for ALS models)
or ON (for Plus and BLS models), and hold the
softkeys for approximately 4 seconds.

The unit powers on in System Configuration mode.

2.

Enter the passcode using the appropriate softkeys.

Note: When you enter System Configuration for

the first time, enter the default pass code of
00000000.

To change a digit, use the Delete softkey to
highlight the entry and re-enter the digit.

When all 8 digits have been entered, the unit
automatically advances to the Select Language
screen.

3.

Use the Prev Value and Next Value softkeys to
toggle the highlighted area between the language
options.

4.

Press the Enter softkey to select the highlighted
language.
All subsequent configuration menus will be in the
selected language.

The unit automatically advances to the Main
Configuration menu.

System Configuration

Set Configuration Passcode

or Power Off to Exit

0 0 0 0 ? ? ? ?

0

3


Delete

2

1

System Configuration

Enter Configuration Passcode

or Power Off to Exi

t

0 0 0 ? ? ? ? ?

0

3


Delete

2

1


Enter

Exit

Config

Next

Value

Prev

Value

System Configuration

Select Language

ENGLISH

FRANCAIS

System Configuration

Print

Change

Config

Exit

Config

Read
From

Card

Restore

Defaults