Nibp settings (optional), Blood pressure units, Trigger nibp meas on bp alarm – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 49: Trigger nibp meas on hr alarm

Advertising
background image

NIBP Settings (Optional)

9650-0902-01 Rev. K

43

NIBP Settings (Optional)

NIBP configuration parameters are available for update only if the NIBP monitoring option is
installed on the unit, and include:

Blood Pressure Units

Trigger NIBP Meas on BP Alarm

Trigger NIBP Meas on HR Alarm

NIBP Auto Interval Default

Adult Initial Cuff Inflation

Pediatric Initial Cuff Inflation

Neonatal Initial Cuff Inflation

NIBP Patient Mode at Power Up

NIBP Button Press & Hold

Enable Motion Tolerance

Blood Pressure Units

This parameter allows you to set the NIBP unit of measurement.

Values are as follows:

Trigger NIBP Meas on BP Alarm

This parameter allows you to specify whether or not the R Series unit initiates a single
additional blood pressure measurement when any of the NIBP alarms trigger.

Values are as follows:

Trigger NIBP Meas on HR Alarm

This parameter allows you to specify whether or not the R Series unit initiates a single blood
pressure measurement when the heart rate/pulse rate alarm triggers.

Values are as follows:

Possible Values:

mmHg (millimeters of mercury), kPa (kilopascals)

Default:

mmHg

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

No

Advertising