Gain on Power Up

This setting allows you to select the default size of the displayed ECG signal when you power
on the unit.

Values are as follows:

Enable Leads Off

This setting specifies whether the unit displays the ECG LEADS OFF message and a flat
dashed ECG waveform when the unit detects that ECG leads do not have proper contact with
the patient or the cable is not properly attached to the R Series unit. When set to No, the unit
displays the POOR LEAD CONTACT message and the detected ECG signals when the cable or
ECG leads are not properly attached to the patient or unit. When set to Yes, the unit displays the
ECG LEADS OFF message and a flat dashed ECG waveform when the cable or ECG leads are
not properly connected.

Caution

Use care when interpreting ECG waveforms when the POOR LEAD CONTACT message
displays; the ECG waveform may include a significant amount of artifact. Whenever possible,
re-attach the leads before interpreting the waveform.

Values are as follows:

Note:

When in Pacer mode, the R Series unit functions as if this configuration is set to Yes.

Enable Pacer Detection

This setting allows you to specify whether the unit should detect pacemaker signals from a
patient with an implantable pacemaker and indicate those signals on the display. When you set
this setting to Yes, a pacemaker marker is displayed on the ECG trace whenever the unit detects
implanted pacer stimuli.

You can override this setting during normal operations. This setting simply indicates the default
setting when you power on the unit.

Values are as follows:

Possible Values:

0.5 cm/mV, 1.0 cm/mV, 1.5 cm/mV, and 2.0 cm/mV

Default:

1.0 cm/mV for Manual mode

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

Yes