Chapter 2

Configuration Settings

26

9650-0902-01 Rev. K

Ped Energy Level Shock 2

This setting allows you to select the energy level used for the second pediatric shock delivered
by the defibrillator when the unit is:

in Advisory mode and using OneStep Pediatric electrodes.

in Manual mode, using OneStep Pediatric electrodes and Basic Auto Energy Escalation is
set to Yes

Values are as follows:

Ped Energy Level Shock 3

This setting allows you to select the energy level used for the third and all subsequent shocks
delivered by the defibrillator when the unit is:

in Advisory mode and using OneStep Pediatric electrodes.

in Manual mode, using OneStep Pediatric electrodes and Basic Auto Energy Escalation is
set to Yes.

Values are as follows:

Auto Analyze

This setting allows you to specify whether the unit automatically repeats ECG rhythm analyses
after the delivery of shocks in Advisory mode and works in conjunction with the Stacked
Shocks setting described on page 30. Enabling this setting also enables the occurrence of post-
shock prompting to DO CPR and PRESS ANALYZE. When set to No, which is the Manual
mode default, the unit does not perform analysis automatically (with the exception of Smart
Alarms and the first analysis when you power on the unit in AED mode, if the unit is
configured to analyze when you power on the unit).

When set to Yes and the Stacked Shocks options is set to 3, the unit automatically reanalyzes
the patient’s ECG rhythm after the first and second shocks in each three-shock sequence.
Following delivery of the final shock, the unit stops automatic reanalysis and either displays a
CHECK PULSE prompt, or immediately begins a CPR interval, depending on additional
settings. This function operates only when the Auto Charge in Advisory Mode is set to Yes.

Values are as follows:

Possible Values:

Pediatric: 1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

Pediatric: 70 J

Possible Values:

Pediatric: 1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

Pediatric: 85 J

Possible Values:

Yes, No

Default:

Yes