Pace Settings

9650-0902-01 Rev. K

33

Pace Settings

The Pace settings are attributes that control functions related to performing pacing with an
R Series unit and include:

Pace Rate Setting on Power-Up

Async Softkey in Pace Mode

Pace Rate Setting on Power-Up

This setting allows you to set the default pace rate, pulses per minute (ppm), of the R Series
when you power on the unit.

Values are as follows:

Async Softkey in Pace Mode

This setting allows you to set the unit to display the Async Pacing On/Off softkey when the
unit is in Pace mode. For information about Asynchronous Pacing, refer to the R Series
Operator’s Guide.

Values are as follows:

Possible Values:

30 ppm to 102 ppm (inclusive in increments of 4 ppm)

Default:

70 ppm

Possible Values:

Yes, No

Default:

Yes