Enable voice in advisory mode, Display do cpr message – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 34

Advertising
background image

Chapter 2

Configuration Settings

28

9650-0902-01 Rev. K

Enable Voice in Advisory Mode

This setting allows you to enable or disable voice prompts when using shock advisory functions
(ECG analysis) in Manual mode. If enabled, the voice prompts sound at the appropriate time. If
disabled, the unit emits two beeps instead of the voice prompts in Manual mode.

Voice prompts are in the language specified by the Language option.

Values are as follows:

Display DO CPR Message

This setting allows you to specify whether the unit displays the configured CPR prompt
(described in CPR Text Message) for the configured CPR time interval after each set of stacked
shocks, even if a perfusing rhythm is present. The unit behaves differently depending on this
particular setting value.

Note:

This setting is always set to Yes in AED mode.

When set to No (the Manual mode default), the unit behaves as follows:

At the end of any analysis with an outcome of no shock advised, the unit displays the NO
SHOCK ADVISED
message. The unit then remains idle for the configured CPR interval
duration. After the CPR interval, if the ANALYZE softkey has not been pressed, the
PRESS ANALYZE message is displayed and remains until a new analysis is started or the
mode is changed.

At the end of a stacked shock sequence (assuming the final shock is delivered) and when
the optional 10 second CHECK PULSE message disappears, the CPR interval begins.
The unit remains idle for the duration of the CPR interval. After the CPR interval, if the
ANALYZE softkey has not been pressed, the PRESS ANALYZE prompt is displayed and
remains until a new analysis is started or the mode is changed.

If the defibrillator disarms itself after it was charged (because the Defib Ready time
elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt
until the ANALYZE softkey is pressed.

When set to Yes, the unit behaves as follows:

At the end of any analysis with an outcome of no shock advised, the unit displays the NO
SHOCK ADVISED
message for 10 seconds, after which the message is cleared and the
configured CPR prompt is displayed for the duration of the CPR interval. After the CPR
interval, the PRESS ANALYZE prompt is displayed until a new analysis is started or a
mode change occurs. You can start a new analysis during the CPR interval by pressing
the ANALYZE softkey.

At the end of a stacked shock sequence (assuming the final shock is delivered) and when
the optional 10 second CHECK PULSE message disappears, the CPR interval begins.
The configured CPR prompt is displayed for the duration of the CPR interval. After the
CPR interval, the CPR prompt is cleared and a new analysis starts automatically.

Possible Values:

Yes/No

Default:

Yes