If the defibrillator disarms itself after it was charged (because the Defib Ready time
elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt
until the ANALYZE softkey is pressed.

Note:

The Display DO CPR Message feature is operational only when the Auto Charge in
Advisory Mode and Auto Analyze options are both set to Yes.

Values are as follows:

CPR Message After No Shock Advised

This parameter determines which CPR message/voice prompt is issued during the CPR interval
after a No Shock Advised analysis result. The selected message is displayed throughout the
entire CPR interval.

Values are as follows:

CPR Message After Shock

This parameter determines which CPR message/voice prompt is issued during the CPR interval
after the delivery of the final shock in a stacked shock sequence. The selected message is
displayed throughout the entire CPR interval.

Values are as follows:

Possible Values:

Yes, No

Default:

Yes

Possible Values:

“If No Pulse, Perform CPR”, “Perform CPR”

Default:

Perform CPR

Possible Values:

“If No Pulse, Perform CPR”, “Perform CPR”

Default:

“Perform CPR”