Stacked shocks, Restart analysis after cpr, Cpr interval – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 36

Advertising
background image

Chapter 2

Configuration Settings

30

9650-0902-01 Rev. K

Stacked Shocks

The setting specifies the number of stacked shocks that must be administered prior to the
activation of a CPR interval and works in conjunction with the Auto Analyze setting. If set to 3,
the unit will auto analyze 3 times before prompting the user. If set to 1, the unit will auto
analyze once.

Values are as follows:

Restart Analysis After CPR

When this parameter is set to Yes, the analysis starts automatically after the CPR interval
following each stacked shock sequence. Analyses do not automatically restart after a CPR
interval following No Shock Advised results. When this parameter is set to No, the PRESS
ANALYZE message is displayed, but the analysis does not restart.

Note:

This setting is always set to Yes in mode.

Values are as follows:

CPR Interval

This setting allows you to specify the duration of the CPR interval as described in the Display
DO CPR Message setting. Values are as follows:

Possible Values:

1, 3

Default:

1

Possible Values:

Yes, No

Default:

No

Possible Values:

1, 1.5, 2, 2.5, 3, 3.5, or 4 minutes

Default:

2 minutes