Extra CPR Interval Before 1st Analysis

This parameter pertains only to AED mode.

When this parameter is set to No (default), and the unit detects a pad connection upon power-
on, it automatically starts an ECG analysis. If the unit is powered on with no pads connected,
the unit displays the ATTACH PADS prompt until pads are connected. Once pads are connected,
the unit automatically starts an ECG analysis.

When this parameter is set to Yes, the unit displays the prompts: CHECK PULSE for 10
seconds, then IF NO PULSE, PERFORM CPR for the configured Extra CPR interval at power
on. At the end of the CPR period, if the unit does not detect a pads connection, it displays the
prompt ATTACH PADS until pads are connected. If a pads connection is detected, the unit
automatically starts an ECG analysis. You can start an ECG analysis during the CPR interval by
pressing the ANALYZE softkey.

Values are as follows:

Duration of Extra CPR Interval

This parameter allows you to specify the duration of the extra CPR interval (see previous
entry). It has no effect if “Extra CPR Interval Before 1st Analysis” is set to No.

Values are as follows:

Display Remaining CPR Timer

This parameter allows you to specify whether to display remaining CPR interval during CPR.

Values are as follows:

Possible Values:

Yes, No

Default:

No

Possible Values:

0.5, 1, 1.5, 2, 2.5, 3, 3.5 or 4 minutes

Default:

2 minutes

Possible Values:

Yes, No

Default:

Yes