Enable check pulse prompt, Check patient prompt, Auto charge in advisory mode – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 33

Advertising
background image

Advisory Defib

9650-0902-01 Rev. K

27

Enable Check Pulse Prompt

This setting works in conjunction with the Auto Analyze function and determines whether the
unit displays the prompt CHECK PULSE for 10 seconds immediately after delivery of the final
shock in a stacked shock sequence, after each No Shock Advised analysis result, or both. After
CHECK PULSE clears, the CPR interval begins.

When set to No, the unit immediately begins the CPR interval following delivery of the final
shock or after each No Shock Advised analysis result.

Values are as follows:

Check Patient Prompt

This setting allows you to change the display message and the voice prompt issued when a
shockable rhythm has been detected by background ECG analysis and Heart Rate Alarms are
enabled.

Values are as follows:

Auto Charge in Advisory Mode

This setting allows you to specify whether the unit automatically charges the defibrillator to the
configured or user-selected energy setting after detection of a shockable rhythm by ECG
analysis. When set to Yes, the unit automatically charges the defibrillator to the configured or
user-selected energy setting. When set to No, it does not.

Values are as follows:

Possible Values:

Yes, No, After No Shock Only, After Shock Only

Default:

No

Possible Values:

CHECK PATIENT, PRESS ANALYZE

Default:

CHECK PATIENT

Possible Values:

Yes, No

Default:

Yes