Ecg settings, Notch filter, Display lead on power-up – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 25: Frequency response

Advertising
background image

ECG Settings

9650-0902-01 Rev. K

19

ECG Settings

ECG monitoring settings are attributes associated with using the R Series unit for ECG
monitoring, and include:

Notch Filter

Display Lead on Power-Up

Frequency Response

Gain on Power Up

Enable Leads Off

Enable Pacer Detection

Notch Filter

This setting allows you to select the ECG Notch Filter frequency for proper ac mains
interference rejection.

Values are as follows:

Display Lead on Power-Up

This setting allows you to select the lead to be displayed when you power on the unit.

Values are as follows:

Frequency Response

This setting allows you to select the ECG filter bandwidth to be used during ECG monitoring.
You can set the unit’s frequency to either help reduce noise, help with the diagnostic response,
or set it as a default response.

Values are as follows:

Possible Values:

50 Hz, 60 Hz

Default:

60 Hz

Possible Values:

Paddles/Pads, Lead I, Lead II, Lead III

Default:

Paddles/Pads

Possible Values:

0.5 - 21 Hz, 0.5 - 40 Hz, 1.0 - 21 Hz

Default:

0.5 - 21 Hz