CPR Settings

9650-0902-01 Rev. K

31

CPR Settings

The CPR settings control functions related to Real CPR Help

®

function:

Enable CPR Voice Prompts

Enable CPR Display

Enable CPR Metronome

Target CPR Depth

Display Fully Release Prompt

Enable CPR Voice Prompts

This setting allows you to enable (Yes) or disable (No) voice prompts during CPR. Voice
prompts are in the language specified by the Language option.

Note:

This setting is always set to Yes in AED mode.

Values are as follows:

Enable CPR Display

This setting allows you to enable (Yes) or disable (No) the display of the CPR indicator on the
monitor.

Values are as follows:

Enable CPR Metronome

This setting allows you to enable (Yes) or disable (No) the audible CPR metronome.

Note:

This setting is always set to Yes in AED mode.

Values are as follows:

Target CPR Depth

This parameter allows you to select the minimum compression depth required for Good
Compressions: 1.5 inches (3.8 cm) or 2.0 inches (5.0 cm). If this parameter is set to 1.5 inches,
the references markers on the CPR bar graph are displayed at 1.5 and 2.0 inches. If this

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

Yes