Chapter 2

Configuration Settings

64

9650-0902-01 Rev. K

Values are as follows:

Procedure

Follow these steps to set up an server IP address.

Possible Values:

XXX.XXX.XXX.XXX where XXX is 0 to 255

Default:

000.000.000.000

Step

Action

1.

From the Select Configuration Categories screen,
select Wi-Fi, then press the Enter softkey.

The Wi-Fi Settings screen appears.

2.

Press the Next Item softkey until Server IP (Defib
History) appears and press the Change Value
softkey.

The Set Defib History Server IP Address screen
appears.

3.

Use the softkey as follows:



— moves the highlight up, down,

left, and right through the character selection
table.

Enter Char — enters the highlighted
character for each digit of the server IP.

Return — returns you to the Wi-Fi
Configuration screen. Press the Return
softkey when you are finished entering the
server IP.

Next

Item

Prev

Item

System Configuration

Wi-Fi Configuration

Defib History

Transfer Mode

DHCP/

Static IP

Both

Static IP

Main

Menu

Exit

Config

Category

List

Change

Value

Full Disclosure

Transfer Mode

Server

Page 1 of 7

System Configuration

Set Defib History Server IP Address

192.168.042.001

Return

Enter

Char

1

Cancel

4
7

2
5
8

3
6
9

0

<

<

.