Basic Defibrillation Settings

The Basic Defibrillation settings control functions related to performing defibrillation with the
R Series unit in Manual mode and include:

Defib Default to Pads

Basic Auto Energy Escalation

Retain Sync After Defib

Defib Ready Hold Time

Enable Remote Sync

Defib Default to Pads

This setting allows you to override the “Display Lead on Power-Up” setting and use Pads/
Paddles whenever the unit is switched to DEFIB mode.

Values are as follows:

Basic Auto Energy Escalation

This setting allows you to configure the unit to automatically increment defibrillation energy to
pre-specified levels after each of the first two shocks in Manual mode (see “Energy Level:
Shock 1” on page 24, “Energy Level: Shock 2” on page 25, and “Energy Level: Shock 3” on
page 25). When set to Yes, the ENERGY INCREMENTED message is displayed whenever the
unit automatically increments the selected defibrillator energy after delivery of a shock with
paddles or MFE pads.

Note:

Manually changing the energy level outside the pre-programmed sequence and
delivering a shock disables the auto energy escalation function until the unit is turned
off for more than 10 seconds, then turned back on.

This option increments the energy level regardless of Shock Advisory activation or status.

Values are as follows:

Retain Sync After Defib

This setting allows you to configure the unit to remain in SYNC mode after a synchronized
cardioversion shock. SYNC mode remains on until the Sync On/Off softkey is pressed again or
the unit is switched out of DEFIB mode.

Values are as follows:

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

No