Alarms Settings
9650-0902-01 Rev. K
37
Default:
Heart Rate:
High: 150 BPM
Low: 30 BPM
Optional Devices:
SpO
2
:
High: 100%
Low: 85%
EtCO
2
:
High: 55 mmHg
7.3 kPa
7.2%
Low: 25 mmHg
3.3 kPa
3.2%
Resp Rate:
High: 120
Low: 5
NIBP Sys:
Adult:
High: 160 mmHg
21.3 kPa
Low: 90 mmHg
12.0 kPa
Pediatric:
High: 145 mmHg
19.3 kPa
Low: 75 mmHg
10.0 kPa
Neonatal:
High: 100 mmHg
13.3 kPa
Low: 50 mmHg
6.7 kPa