Alarm limits at power up – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 41

Advertising
background image

Alarms Settings

9650-0902-01 Rev. K

35

Alarm Limits at Power Up

This option allows you to enable or disable, and set high and low threshold values for the ECG
Heart Rate, and the SpO

2

alarms on the R Series unit. All alarms listed below (Heart Rate and

optional devices SpO

2

, EtCO

2

, NIBP) are enabled by default. Values are as follows:

Possible Values:

Heart Rate:

High: 60 - 280 beats per minute (BPM) inclusive in 5 BPM increments

Low: 20 - 100 beats per minute (BPM) inclusive in 5 BPM increments

Optional Devices:

SpO

2

:

High: 72-100% in 1% increments

Low: 70-98% in 1% increments

EtCO

2

:

High: 5-100 mmHg in increments of 5

0.6-13.3 kPa in increments of .67

0.6-13.1% in increments of .66

Low: 0-95 mmHg in increments of 5

0.0-12.6 kPa in increments of .67

0-12.5% in increments of 0.66

Resp Rate:

High: 5-150 breaths per minute in increments of 5

Low: 0-100 breaths per minute in increments of 5

NIBP Sys:

Adult:

High: 80-260 mmHg in increments of 5

10.7-34.7 kPa in increments of 0.67

Low: 40-140 mmHg in increments of 5

5.3-18.7 kPa in increments of 0.67

Pediatric:

High: 80-160 mmHg in increments of 5

10.7-21.3 kPa in increments of 0.67

Low: 35-140 mmHg in increments of 5

4.7-18.7 kPa in increments of 0.67

Neonatal:

High: 60-130 mmHg in increments of 5

8.0-17.3 kPa in increments of 0.67

Low: 25-120 mmHg in increments of 5

3.3-16.0 kPa in increments of 0.67