Chapter 2

Configuration Settings

44

9650-0902-01 Rev. K

NIBP Auto Interval Default

This parameter allows you to set the default for the amount of time between automatic
measurements.

Values are as follows:

Adult Initial Cuff Inflation

This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which
the NIBP cuff inflates at the beginning of its first measurement after power up).

Values are as follows:

Pediatric Initial Cuff Inflation

This parameter allows you to set the default pediatric NIBP cuff inflation pressure (the pressure
to which the NIBP cuff inflates at the beginning of its first measurement after power up, if set to
pediatric).

Values are as follows:

Neonatal Initial Cuff Inflation

This parameter allows you to set the default NIBP cuff inflation pressure (the pressure to which
the NIBP cuff inflates at the beginning of its first measurement after power up, if set to
neonatal).

Values are as follows:

Possible Values:

2.5, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 45, 60, 90, 120 minutes

Default:

30 minutes

Possible Values:

mmHg: 120, 140, 160, 180, 200, 220, 240, 260
kPa: 16.0, 18.7, 21.3, 24.0, 26.7, 29.3, 32.0, 34.7

Default:

mmHg: 160

kPa: 21.3

Possible Values:

mmHg: 80, 90, 100, 110, 120, 130, 140, 150
kPa: 10.7, 12.0, 13.3, 14.7, 16.0, 17.3, 18.7, 20.0

Default:

mmHg: 120

kPa: 16.0

Possible Values:

mmHg: 60, 70, 80, 90, 100, 110, 120, 130
kPa: 8.0, 9.3, 10.7, 12.0, 13.3, 14.7, 16.0, 17.3

Default:

mmHg: 90

kPa: 12.0