Alarms active at power up, Procedure, System configuration – ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 45

Advertising
background image

Alarms Settings

9650-0902-01 Rev. K

39

Procedure

Follow these steps to change the state or value of the ECG Heart Rate alarms:

Alarms Active at Power Up

This setting allows you to specify whether alarms are automatically active when you power on
the unit. This setting applies to all alarms enabled by the Alarm Limits At Power Up setting.

Step

Action

1.

From the Select Configuration Categories screen,
select Alarms, then press the Enter softkey.

The Alarms Setting screen appears.

2.

Select Alarm Limits at Power Up, and press the
Change Value softkey.

The Set Alarm Defaults screen appears.

3.

Modify alarms using the softkeys as follows:

Next Param and Prev Param — moves the
highlight through the various physiological
settings.

Next Field — moves highlight to the next field
to the right. Use this key to change the High and
Low settings.

Change Value — allows you to change the
value of the highlighted field. When you press
this softkey, the following softkeys become
available:

Inc > and Dec < — increases and decreases
the value of the field.

Enter — applies the value you specified for
this field.

Cancel — returns you to the previous screen
without applying any changes.

Return enters the values you specified, and
returns you to the previous screen.

Next
Item

Prev
Item

System Configuration

Alarm Settings

Set/Review

Main

Menu

Exit

Config

Category

List

Change

Value

Page 1 of 1

No

ALARMS ACTIVE

AT POWER UP

ALARM LIMITS

AT POWER UP

Set/Review

Change

Value


Return

System Configuration

Set Alarm Defaults

Parameter

State

Low

High

30

150

ECG HR

ENABLE

Next

Param

Prev

Param

Next

Field

85

100

ENABLE

SpO2

25

55

ENABLE

EtCO2

5

120

ENABLE

RESP RATE

90

160

ENABLE

NIBP SYS

50

110

ENABLE

NIBP DIA

60

130

ENABLE

NIBP MEAN

75

145

ENABLE

NIBP p SYS
Page 1 of 2