Chapter 2

Configuration Settings

24

9650-0902-01 Rev. K

Advisory Defib

The Advisory Defibrillation settings control functions related to performing defibrillation with
an R Series unit in Advisory mode and include:

Energy Level: Shock 1

Energy Level: Shock 2

Energy Level: Shock 3

Ped Energy Level Shock 1

Ped Energy Level Shock 2

Ped Energy Level Shock 3

Auto Analyze

Enable Check Pulse Prompt

Check Patient Prompt

Auto Charge in Advisory Mode

Enable Voice in Advisory Mode

Display DO CPR Message

CPR Message After No Shock Advised

CPR Message After Shock

Stacked Shocks

Restart Analysis After CPR

CPR Interval

Energy Level: Shock 1

This setting allows you to select the energy level used for the first shock delivered by the
defibrillator when the unit is:

in AED mode or Advisory mode and using pads

in Manual mode, using external paddles or pads

Energy Level: Shock 1 is the energy level to which the unit defaults when you power on the
unit with external paddles or pads connected.

Values are as follows:

Possible Values:

Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J

Default:

Adult: 120 J