ZOLL R Series Monitor Defibrillator Rev K CONFIGURATION User Manual

Page 42

Advertising
background image

Chapter 2

Configuration Settings

36

9650-0902-01 Rev. K

Possible Values:

NIBP Dia:

Adult:

High: 50-200 mmHg in increments of 5

6.7-26.7 kPa in increments of 0.67

Low: 25-100 mmHg in increments of 5

3.3-13.3 kPa in increments of 0.67

Pediatric:

High: 50-130 mmHg in increments of 5

6.7-17.3 kPa in increments of 0.67

Low: 20-100 mmHg in increments of 5

2.7-13.3 kPa in increments of 0.67

Neonatal:

High: 30-105 mmHg in increments of 5

4.0-14.0 kPa in increments of 0.67

Low: 20-100 mmHg in increments of 5

2.7-13.3 kPa in increments of 0.67

NIBP Mean:

Adult:

High: 60-220 mmHg in increments of 5

8.0-29.3 kPa in increments of 0.67

Low: 30-120 mmHg in increments of 5

4.0-16.0 kPa in increments of 0.67

Pediatric:

High: 60-140 mmHg in increments of 5

8.0-18.7 kPa in increments of 0.67

Low: 30-120 mmHg in increments of 5

4.0-16.0 kPa in increments of 0.67

Neonatal:

High: 35-110 mmHg in increments of 5

4.7-14.7 kPa in increments of 0.67

Low: 30-105 mmHg in increments of 5

4.0-14.0 kPa in increments of 0.67