Chapter 2

Configuration Settings

40

9650-0902-01 Rev. K

Values are as follows:

Display Settings

The Display settings are attributes that control what and how things appear on the R Series
monitor and include:

Display Time

Select Trace 2 At Power Up

Select Trace 3 At Power Up

ECG Color

SpO2 Color

CPR Color

EtCO2 Color

NIBP Color

SPO2 at Power Up

Display Time

This setting specifies whether to display the elapsed time since the unit was turned on (Elapsed
Time), or the real time (Real Time), or not to display the time (Off).

The elapsed time continues to be counted for up to 10 seconds after you power off the unit. This
gives the operator adequate time to change the battery without resetting the elapsed timer.

Elapsed time is reset to zero (00:00) whenever the unit has been off for more than 10 seconds.

Values are as follows:

Select Trace 2 At Power Up

This setting allows you to set which waveform will appear in the Trace 2 position on the display
when you turn on the R Series unit.

Values are as follows:

Note:

When you set this setting to 3 Lead ECG, the Select Trace 3 At Power Up setting
automatically defaults to 3 Lead ECG.

Possible Values:

Yes, No

Default:

No

Possible Values:

Real Time, Off, Elapsed Time

Default:

Off

Possible Values:

Off, 3 Lead ECG, CPR, Filtered ECG, SpO2, EtCO2

Default:

Filtered ECG