Advisory Defib

9650-0902-01 Rev. K

25

Energy Level: Shock 2

This setting allows you to select the energy level used for the second shock delivered by the
defibrillator when the unit is:

in AED mode or Advisory mode and using pads

in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes

Values are as follows:

Energy Level: Shock 3

This setting allows you to select the energy level used for the third and all subsequent shocks
delivered by the defibrillator when the unit is:

in AED mode or Advisory mode and using pads

in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes

Values are as follows:

Ped Energy Level Shock 1

This setting allows you to select the energy level used for the first pediatric shock delivered by
the defibrillator when the unit is:

in Advisory mode and using OneStep™ Pediatric electrodes.

in Manual mode, using OneStep Pediatric electrodes

Ped Energy Level: Shock 1 is the energy level to which the unit defaults when you power on the
unit with pediatric electrode pads.

Values are as follows:

Warning!

The R Series will default to pediatric Shock 1, 2, and 3 energies only when ZOLL

®

OneStep pediatric electrodes are used. With pedi-padz

®

and pedi-padz II (or other non-

ZOLL pediatric electrodes), you are required to manually set the selected energy to
pediatric levels, per institutional protocols.

Possible Values:

Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J

Default:

Adult: 150 J

Possible Values:

Adult: 1-10, 15, 20, 30, 50, 75, 100, 120, 150, and 200 J

Default:

Adult: 200 J

Possible Values:

Pediatric: 1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

Pediatric: 50 J