• Rate of Change Increase

Select the number of data samples that determines the interval of the rate of change

of an upper limit alarm between [1] and [15].

• Rate of Change Decrease

Select the number of data samples that determines the interval of the rate of change

of an lower limit alarm between [1] and [15].

• Alarm Hysteresis

Set the alarm hysteresis to [ON] or [OFF]. When it is [ON], the hysteresis is set to 0.5%

of the scale or the measurement span.

Remote (Opton)

You can assign items to be controlled by the eight remote control terminals. This is

possible, if the measurement remote input is available. For details, refer to the CX

User’s Manual (IM 04L31A01-01E or IM 04L31A01-03E).

Select a remote number, and then click the [Copy] button. This copies the setting of that

remote number. When you want to paste the copied setting, select the remote number

where you want to paste the setting, and then click the [Paste] button.

Scan Interval/Memory

Set these parameters when the data type is [EVENT & DISP] or [EVENT].

Click here. (Or choose the [Setting] menu - [SETUP [Basic] Setting] - [Setting].)

Select the channels that

you want to sample.

A/D Integrate

Select from [Auto], [50 Hz], [60 Hz], and [100 ms]. The [Auto] setting automatically

detects the CX power supply frequency, and switches the integration time.

Scan Interval

Select [1s] or [2s].

8.7 Basc Measurement Functon Basc Settngs