115 • Logs Button

7.

In the graph part of the screen,

You can choose which gas(es) you want to graph by selecting or
deselecting the boxes next to each gas name.

You can select the display units for the catalytic LEL channel (%LEL, ppm,
vol%). While readings can be recorded in any of these units and stored in
the same file, those readings may only be graphed in one unit at a time. If
%LEL is chosen as the unit, any ppm or %volume data is converted and
graphed as %LEL. The user may change back and forth between the units
for the graph and still have all data shown. The CAT Range Select box is
where the user may select %LEL, ppm, or %volume as the units for the
graph.

If installed, an IR HC or CH

4

channel can be graphed in %LEL or % volume

in order to accurately depict the full range of the sensor.

If the channel is configured as a %LEL only channel, then data will only be
recorded in %LEL and the range will be 0-100% LEL. Any gas
concentration above that level will not be recorded.

If the channel is configured as a %LEL/% volume autoranging channel,
then data will be recorded in %LEL and % volume where appropriate. It will
record data in %LEL up to 100% LEL and then it will start recording data in
% volume up to 100% volume. If %LEL is chosen as the graphing units, all
data collected for a %LEL only channel will be displayed. Any data
collected in an autoranging configuration that exceeds 100% LEL will not
be shown. In order to view data above 100% LEL, % volume must be
selected. When % volume is selected, all data will be shown in terms of %
volume.