RKI Instruments SDM-E2 User Manual

Page 134

Advertising
background image

129 • Logs Button

%LEL is chosen as the unit, any ppm or %volume data is converted and
graphed as %LEL. The user may change back and forth between the units
for the graph and still have all data shown. The CAT Range Select box is
where the user may select %LEL, ppm, or %volume as the units for the
graph.

If installed, an IR HC or CH

4

channel can be graphed in %LEL or % volume

in order to accurately depict the full range of the sensor.

If the channel is configured as a %LEL only channel, then data will only be
recorded in %LEL and the range will be 0-100% LEL. Any gas
concentration above that level will not be recorded.

If the channel is configured as a %LEL/% volume autoranging channel,
then data will be recorded in %LEL and % volume where appropriate. It will
record data in %LEL up to 100% LEL and then it will start recording data in
% volume up to 100% volume. If %LEL is chosen as the graphing units, all
data collected for a %LEL only channel will be displayed. Any data
collected in an autoranging configuration that exceeds 100% LEL will not
be shown. In order to view data above 100% LEL, % volume must be
selected. When % volume is selected, all data will be shown in terms of %
volume.