Instrument Function Menu • 154

Conversion Table Tab

The Conversion Table tab is used to view the pre-defined relative response gases for
the catalytic combustible channel and to edit or add user-defined gases. The pre-
defined gases can be found in the Pre-Defined Table tab while the user-defined
gases can be found under the User-Defined Table tab.

There are 8 columns in both the Pre-Defined Table tab and the User-Defined Table
tab:

No
This column represents the gas number. The gas numbers are 1-30.

Name
This is what will appear in the Catalytic Sensor Relative Response Screen in the
EAGLE 2. The name can be up to 3 characters long and the characters must be
upper case letters or numbers. No special characters may be used in the Name
column.

Long Name
The Long Name column is used to better describe the target gas. It may contain
any character in upper- or lower-case.

Factor
This value is the response factor for the listed gas relative to methane. The

Figure 117: Conversion Table Tab, Pre-Defined