RKI Instruments SDM-E2 User Manual
Page 213

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 208
If you are calibrating single channel special sensor instruments and have a
manifold assembled, the gas order confirmation screen will appear as shown
below. One sensor will be tested using GAS 1 (in this case, NH
3
). The
remaining sensors will be tested using GAS 2. The GAS 2 shown in the
column directly to the right of the GAS 1 column (in this case, IBL) will always
be calibrated before the GAS 2 shown in the following columns. The GAS 2
selection boxes are deselected by default. You must select the GAS 2 boxes if
you wish to calibrate the special sensors. If you do not wish to calibrate the
special sensors, leave the GAS 2 boxes deselected. Be sure that the
appropriate GAS 2 cylinder is connected to the GAS 2 fitting. The PC
Controller program will prompt you to change the GAS 2 cylinder during the
calibration.
Figure 149: Gas Order Confirmation Screen, Single Channel Units, Manifold