RKI Instruments SDM-E2 User Manual

Page 83

Advertising
background image

Bump Testing an Instrument with Special Sensors • 78

12. After air is applied, and if all channels passed the zero adjustment, calibration

gas from the GAS 1 cylinder will be applied. The GAS 1 indicator on the
program screen will come on and calibration gas will be applied to the
instrument(s) for the duration specified in the Gas Exposure Time parameter in
the Configuration Window.

Figure 57: GAS 1 Flowing

The PC Controller Program will analyze the results. If the bump test on the
standard 4 sensors fails and Force Calibration is selected, a calibration will
automatically begin and GAS 1 calibration gas will continue to flow.

Advertising