RKI Instruments SDM-E2 User Manual

Page 37

Advertising
background image

Setting Up the PC Controller Program • 32

parameters in connected instruments. See “Edit Function” on page 140 for
a complete description of the Edit function. If it is not selected, the Edit
function is not active. Select or deselect the Change Parameter box.

Auto Power OFF Time

This is the length of time that will pass after the last operation is finished
before the program will automatically shut off the EAGLE 2. If a successful
bump test or calibration is performed, the program will automatically shut
off the EAGLE 2 in 20 seconds regardless of the Auto Power OFF Time
setting. Use the arrows to the right of the current setting to change the
value or highlight the current setting and type in the desired value. It is
defined in seconds with a maximum setting of 3600 seconds (1 hour). The
factory setting is 3600 seconds.

6.

Select the Data Logger options.

Down Load Data Logger

The Down Load Data Logger selection allows you to have saved data
automatically downloaded from the EAGLE 2 once it is connected to the
PC Controller program. The data will automatically be downloaded based
on the frequency set in the Down Load Interval option. The factory setting
is deselected. Select or deselect the Down Load Data Logger box.

Clear After Down Load

The Clear After Down Load selection allows you to automatically clear the
EAGLE 2 data logger after an automatic download. The factory setting is
deselected. Select or deselect the Clear After Down Load box. If Down
Load Data Logger is not selected, selecting Clear After Down Load has no
effect.

Down Load Interval

The Down Load Interval is the length of time, in days, that will pass before
the PC Controller program will automatically download data from an
EAGLE 2 if Down Load Data Logger is selected. If the EAGLE 2 is
connected to the program before the interval has passed, a data download
will not automatically occur. The Down Load Interval can be set anywhere
from 1 to 60 days. The factory setting is 1 day. Use the arrows to the right
of the current setting to change the interval or highlight the current setting
and type in the desired value.

7.

SDM Selection

If you select the Initialize Display Order selection box, the PC Controller
program will reset the numbers it has assigned to specific docking stations and
reassign them as you turn SDM-E2 docking stations on and connect them to
the program. See “Setting Up the SDM-E2 Display Order” on page 35 for
instructions to use this feature.