Logs Button • 120

Calibration Report control button along the top of the Logs window. A dialog
box will appear confirming if you want to print. Click OK.

10. To print detailed calibration and bump test data (if available) for each

calibration and bump test file for a specific serial number, be sure that the
desired serial number is selected in the upper left frame and click the Bump
Test & Calibration Report
control button along the top of the Logs window. A
dialog box will appear confirming if you want to print. Click OK.

11. To save calibration data from a specific date folder, be sure the desired date

folder is highlighted and click the Save control button in the upper left corner of
the Logs window. A “Save As” dialog box will appear for you to specify the file
name and file location.

Figure 89: Save As Dialog Box

The file type will be “.csv” (comma separated values).

Specify the file name and location and click Save.

12. To copy calibration data in a date folder to the clipboard, be sure the desired

date folder is selected and click the Save to Clipboard control button. The
data will be saved to the clipboard. It can then be pasted into a document by
using the Paste command in an application.

Event Data

The EAGLE 2 saves the 100 most recent alarm events and the 100 most recent
trouble events. When an EAGLE 2 is connected to the PC Controller Program and a
complete download is performed, the alarm and trouble event files are retrieved and