Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 196
Figure 138: Second GAS 2 Flowing
The PC Controller Program will analyze the results. If the bump test on the
second special sensor fails and Force Calibration is selected, a calibration will
automatically begin on that sensor and GAS 2 calibration gas will continue to
flow.
15. Repeat step 14 for the remaining special sensors installed in the instruments.