Instrument Function Menu • 144

choose to continue accumulating TWA and PEAK readings and the time in
operation from the last time the EAGLE 2 was used or start collecting new TWA
and PEAK readings and reset the time in operation. With Lunch Break not
selected, the EAGLE 2 automatically starts new TWA and PEAK reading
collection and resets the time in operation at startup.

Data Log Overwrite: With Data Log Overwrite selected (factory setting), the
EAGLE 2 writes over the oldest data with new data when the data logger
memory is full. With Data Log Overwrite not selected, the EAGLE 2 stops saving
data to the data logger when the data logger memory is full.

Data Log Memory: With Data Log Memory selected (factory setting), the Data
Logging Screen in Display Mode asks whether you want to clear the logged data
in addition to showing the log time remaining. With Data Log Memory not
selected, the Data Logging Screen only shows the remaining log time and does
not give you the opportunity to clear the logged data.

Password Protection: With Password Protection selected, the EAGLE 2 prompts
you for a password when you enter Calibration Mode or Setup Mode. With
Password Protection not selected (factory setting), no password is required to
enter Calibration Mode or Setup Mode. If you choose to select Password
Protection, delete the current password in the box below the selection and enter
the desired password.