217 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations

If at any point during the calibration the flow rate to the instrument becomes
too low, the PC Controller will abort the calibration and display a failed
calibration screen. The CAL

S

LED on the SDM-E2 control panel will be

solid red. Click OK to return to the main program window or double click the
instrument information line to view the calibration results in the Logs
window. The CAL

S

LED on the SDM-E2 control panel will turn off.

Confirm all tubing connections are correct and that all lines are clear.

Figure 157: Low Flow Calibration Failure

17. For all calibration result outcomes, the instruments will be displayed in the

main program window in Details View. To return to Icons View, click the Icons
radio button.

18. All calibration results can be viewed in the Logs window. See “Calibration

Data” on page 118 for instructions for viewing calibration data.