193 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-

12. After air is applied, and if all channels passed the zero adjustment, calibration

gas from the GAS 1 cylinder will be applied. The GAS 1 indicator on the
program screen will come on and calibration gas will be applied to the
instruments for the duration specified in the Gas Exposure Time parameter in
the Configuration Window.

Figure 136: GAS 1 Flowing

The PC Controller Program will analyze the results. If the bump test on the
standard 4 sensors fails and Force Calibration is selected, a calibration will
automatically begin on the standard 4 sensors and GAS 1 calibration gas will
continue to flow.